Replace the electronic weighing instrument, replacing the old instrument, a new instrument, that is to be noted here that a sensor input addresses and associated parameters, the correct procedure is:
1, modify the meter value and the maximum capacity;
2, addresses input sensors and related parameters and to ensure the switching;
3, adjustment of the instrument of serial port data format and the baud rate.
4, restarting the instrument, observation instrument and zero point of the eccentric load test and a weighing scale body, after the resumption of normal use.
The above - mentioned step has definite logical relationship, not to reverse is performed.According to experience, if the previous modifications to the address input sensors and related parameters, after modifying the index value and a maximum weighing capacity, the gauge display will deviate significantly from zero, if a sensor input parameter is not correct, the scale of the eccentric load is definitely not normal.